Qingdao Puren Instrument Co., Ltd

PRODUCTS

Online ion chromatograph, dual-system ion chromatograph, portable ion chromatograph, atmospheric aerosol on-line detector and related accessories.

Product Details

Features:

Integrated system with double work efficiency:

Dual lC system contains two conductivity detectors、two pumps、two valves、two columns and only one work station into one integrated structure, which effectively overcomes mutual interference between two systems. It is high work efficiency, cost-effective and space-saving.
Mono IC system: With effective column/valve switch mechanism, it could achieve time-sharing detection of anions and cations.


Powerful functional extensions:

Gradient elution can be done with the eluent generator or extra pump/valve mechanism.
Voltammetric detector is optional to detect bromide, iodide, sulfide, cyanide, lead, copper, zinc, cadmium, etc.
Auto-sampler is optional which could operate 100 sample injections without manual duty.


Technical specifications:

PIC-10A Ion Chromatograph (conductivity detector)

Detection limit (injection volume 25μL: Cl-≤0.0005mg/L (hydroxide eluent); BrO3-≤0.0005 mg/L(hydroxide eluent); Li+≤0.002mg/L; Linear range: ≥103(tested by Cl-); correlation coefficient: ≥0.999; Baseline noise: ≤0.5%FS; Baseline drift: ≤1.0FS; Repeatability for qualification: ≤0.5%; Repeatability for quantification: ≤1.0%; Max pressure for pump: ≥36.0MPa; Thermostatic stability: ≤0.1℃/h


PIC-10 Ion Chromatograph

Conductivity detector:  

Detection limit (injection volume 25μL): Cl-≤0.005mg/L (carbonate eluent); Br03-≤0.005 mg/L (carbonateeluent); Li+≤0.005mg/L; Linear range: ≥103(tested by Cl-); correlation coefficient: ≥0.999; Baseline noise:≤0.5%FS; Baseline drift: ≤1.5FS

Amperometric detector:

Detection limit (injection volume 25μL): CN-≤0.001mgL, I-≤0.003mg/L,Cd2+≤0.0001mg/L; correlation coefficient: ≥0.999;
Max pressure for pump: ≥36.0MPa; Thermostatic stability: ≤0.1C/h